Probe only:
Dimensions: measuring chamber: approx. 2.3 cm high, Ø 2 mm, total contact surface: Ø 4 mm, length of the probe: approx. 15.3 cm, cable length: 1.3 m, weight: approx. 75 g (including cable), Resolution: relative humidity ± 0.01 % RH, Temp.: ± 0.01 °C, TEWL: 0.1 g/h/m², Measuring area: TEWL 0-320 g/h/m²
Operation conditions: T: 10-40° C, RH: 30-70 % RH
Measurement uncertainty:
TEWL: ±10 % of the measured value plus ±1 g/h/m² for RH ≤ 80 %, ± 20 % of the measured value plus ± 2 g/h/m² for RH > 80 %
Temperature (T): ± 0.1° C for 20 °C to 60 °C and ±0.2 °C outside this range
Rel. humidity (RH): ± 2.0 % for 0 to 80 % RH and 10 °C to 50 °C, ± 3.0 % outside these ranges
Technical changes may be made without prior notice.